Multidimension Liquid Chromatography

 











 

Fluorescence Detector

 

The Fluorescence Detector aids in selectivity and can have a better detection limit for many compounds. Fluorescence is the property of some chemicals to absorb light and emit it nearly instantly at a longer wavelength. This detector has four options for excitation and one emission or vice versa.

Performance 10 fg anthracene LOD, Ex 250nm, Em 400 400nm H2O>400 Raman, Ex 350nm, Em 397nm, dark value 450nm, standard flow cell
Light Source 20W Xenon flash lamp
normal mode: 20W
economy mode: 5W
Pulse Frequency 365Hz for signal mode
73 Hz for economy mode
Excitation Monochromator Concave holographic grating, F/1.6, blaze 300nm, range 200-700nm and zero-order, bandwidth 20nm
Photomultiplier 200-600nm, exchangeable
Reference system Inline excitation measurement
Time Programming Up to four signals, response time, PMT gain, baseline behavior (append, free, zero), spectral parameters
Spectral Acquisition Excitation or emission spectra, scan speed 28ms per datapoint
Step Size 1-20nm
Spectra Storage All
Wavelength repeatability ±0.2nm
Wavelength Accuracy ±3nm
Flow Cells Standard: 8µl volume, 20 bar (2MPa) pressure maximum, quartz
Environment 0-40°C constant temperature at <95% humidity (non-condensing)