Multidimension Liquid Chromatography
![]() ![]() ![]() ![]() |
![]() ![]() |
|---|
Fluorescence Detector

The Fluorescence Detector aids in selectivity and can have a better detection limit for many compounds. Fluorescence is the property of some chemicals to absorb light and emit it nearly instantly at a longer wavelength. This detector has four options for excitation and one emission or vice versa.
| Performance | 10 fg anthracene LOD, Ex 250nm, Em 400 400nm H2O>400 Raman, Ex 350nm, Em 397nm, dark value 450nm, standard flow cell |
| Light Source | 20W Xenon flash lamp normal mode: 20W economy mode: 5W |
| Pulse Frequency | 365Hz for signal mode 73 Hz for economy mode |
| Excitation Monochromator | Concave holographic grating, F/1.6, blaze 300nm, range 200-700nm and zero-order, bandwidth 20nm |
| Photomultiplier | 200-600nm, exchangeable |
| Reference system | Inline excitation measurement |
| Time Programming | Up to four signals, response time, PMT gain, baseline behavior (append, free, zero), spectral parameters |
| Spectral Acquisition | Excitation or emission spectra, scan speed 28ms per datapoint |
| Step Size | 1-20nm |
| Spectra Storage | All |
| Wavelength repeatability | ±0.2nm |
| Wavelength Accuracy | ±3nm |
| Flow Cells | Standard: 8µl volume, 20 bar (2MPa) pressure maximum, quartz |
| Environment | 0-40°C constant temperature at <95% humidity (non-condensing) |